哈希2100N浊度计-E10报错维修记
仪器信息网·2025-10-08 03:58

Core Viewpoint - The article discusses the repair process of the Hach 2100N Turbidimeter, highlighting the troubleshooting steps taken to resolve an error and the importance of understanding electronic components in maintenance [3][8][16]. Instrument Configuration - The Hach 2100N Turbidimeter has a measurement range typically from 0 to 1000 NTU, with some versions reaching up to 4000 NTU. The highest accuracy is achieved in the low range of 0-10 NTU [5]. - The instrument features a resolution of 0.01 NTU in low range, 0.1 NTU in the 10-100 NTU range, and 1 NTU in high range. It uses a tungsten lamp as a light source, which differs from newer models that utilize LED technology [6]. Accuracy and Issues Encountered - The accuracy of the instrument is ± 2% of the reading or ± 0.01 NTU, whichever is greater, when using StablCal standard solutions within the 0-1000 NTU range [7]. - An error message "Err10" was displayed upon startup, prompting a review of the instrument's manual for troubleshooting [8]. Troubleshooting Process - Upon inspection, it was found that the lamp was not lit, leading to the hypothesis that there might be an issue with the lamp control circuit. A multimeter was used to check the circuit continuity, revealing an abnormality in a CR4 diode [9]. - The CR4 diode was identified as a switching diode related to the lamp control. After sourcing a replacement from inventory, the faulty diode was replaced, restoring normal operation of the instrument [14]. Summary and Considerations - The repair involved addressing a circuit board issue due to a faulty switching diode, which, once replaced, allowed the instrument to function correctly. This experience underscored the importance of practical knowledge in electronics and the potential for cost savings in repairs [16]. - Caution is advised in circuit board repairs due to the high level of expertise required, as improper handling could lead to more significant issues and increased repair costs [16].